B2B
About this product

Product advantages

1. The software: PID calculation is used to calculate accurately control the cooling capacity, making the temperature stable, and applying different loads and their heat generation within the capacity range of the refrigeration system.
2. The refrigeration system adopts cold system (R404 and R23) and adopts electronic expansion valve for energy-saving control to reduce energy consumption (about 40%~50%).
3. Temperature and humidity control range is wide (-70~180℃ & 10%~98%R.H)and accurate control(0.01℃).
4. Mitsubishi PLC + industrial-class man-machine interface, equipment with strong anti-interference, the system is stable and reliable, and low failure rate. After the fault occurs, the equipment alarm is the corresponding processing method of intelligent prompt fault, automatically record the equipment operation data and fault information to help customers to quickly troubleshoot faults and restore the equipment normal operation.

Product details

  • Application:

Uses to test material structure or composite material, testing product reliability in extreme high temperature to very low temperature in a moment, which to test the chemical changes or physical damage caused by it in the most time. Suitable objects include metal, plastic. Rubber, electronics and other materials, it can be used as product improvement base or reference.

  • Technical advantages

1. Adopted temperature and humidity controller with independent intellectual property right.

2. Adopted Mitsubishi PLC+ industrial-level man-machine interface, with strong anti-interference, stable system and low failure rate.

3. Man-machine interactive output operation, recording equipment operation data and fault information automatically. 

4. Historical data of temperature and humidity is exported by U disk, and also EXCEL format table or BMP format image output can be selected with easy operation. 

5. Intelligent prompt corresponding processing method for fault.

6. Adopted PID+ fuzzy control method to save energy and wide range of temperature and humidity control.

7. Wide range of temperature and humidity control.

8. Refrigeration system adopts cold system (R404 and R23). Adopted electronic expansion valve energy saving control, through the software automatic adjustment of valve opening to achieve the stability of temperature, low temperature (below 0℃) stability process heater does not work, the compressor with refrigeration flow reduced power consumption correspondingly reduced, to achieve the purpose of energy saving (about 40~50%).

  • Safety features 
1. Test sample overtemperature protection function.

2. Circulation motor overload protection function.

3. Compressor overload protection function.

4. Compressor overpressure function.

5. Heater/humidifier air burn protection function.

6. Water shortage protection.

7. Input power supply missing phase, reverse phase protection function.

8. Lighting lamp / wire anti-condensation/no fuse independent protection switch in the secondary circuit.

9. Electric leakage protection.

10. Three color light alarm function.

11. Test sample terminal and equipment chain control protection.

12. Equipment fault shutdown, the cycle fan delay stop turn (delay can be set, 0 to 50min, default 5min).

13. Non-metallic materials such as wire / cable / terminals / insulation materials / sound-absorbing materials (if sometimes) are made of flame retardant materials. 

  • Chamber structure and materials
1. Heat preservation layer: Polyurethane foaming + high temperature rock wool, the (left, right side) thickness150mm, the (top, bottom side) thickness100mm.

2. Inner cabin: the inner cabin adopts thickness 1.0mm SUS# 304 heat and cold resistant stainless steel plate, external chamber adopts 1.0 mm plastic coating or color steel plate. It is fully sealed and welded, no water seepage when high temperature and high humidity, Integrated stamping and forming of humidified water plate.

3. Control cabinet and panel: at right side, adopts electrolytic plate plastic spraying, color is Simplewell standard color; Equipped with touch-type man-machine dialog interface, leakage protection switch, emergency stop button, power switch, overtemperature protector, USB data exchange interface, etc. 

4. Doors: electrolytic plate with plastic coating, left single door; size: W 800mm×H 840mm.

5. Base, bottom plate:base frame is welded with steel plate, and bottom plate is reinforced with 3mm steel plate.

6.  Glass window: 1 piece of 3 layers of hollow tempered glass is installed on door, visible range: W 300mm x H 450mm; With electric heating (automatic regulation) anti-frosting, anti-condensation device.

7. Lighting: Lamp voltage is AC12V to prevent the risk of tester electric shock once lamp fault; Automatic delay turned off time after opened.

8. Test hole: 1 piece of φ50mm test hole (located on the left side of chamber, φ100mm test hole and quantity is optional).

9. Pressure balance port: mechanical passive, maintenance-free.

10. Sample shelf: 2 layers round stainless steel bar sample frame and interval is adjustable, bearing weight (uniform): 25kg/ layer.

  • Comply to the standards:

IEC 60068-2-1:2007 Environmental testing –Part 2:Tests – Tests A: Cold;

IEC 60068-2-2:2007 Environmental testing - Part 2-2: Tests - Test B: Dry heat;

IEC60068-2-78:2012 Environmental testing - Part 2-78: Tests - Test Cab: Damp heat, steady state;

IEC 60068-2-30:2005 Environmental testing - Part 2-30: Tests – Test Db: Damp heat, cyclic.

  • Technical parameters

Model

STH150

STH225

STH408

STH1000

Volume(L)

150

225

408

1000

Temperature range(℃)

Min

-20

-40

-70

-20

-40

-70

-20

-40

-70

-20

-40

-70

Max

150(+180℃ optional)

Humidity range

20%RH~98%RH(at temperature 20℃~85℃)

Temperature fluctuation

≤±0.5℃

Humidity fluctuation

≤±3%RH

Temperature deviation

±2.0℃

Humidity deviation

±3.0%RH(>75%): ±5.0%RH(≤75%)

Heating-up time

20℃~+150℃

≤50min

Cooling down time

+20℃~-20℃   Cooling down time ≤45,min

+20℃~-40℃   Cooling down time≤70,min

+20℃~-70℃  Cooling down time≤90,min

Inner size (mm)

W

500

500

600

1000

H

600

750

850

1000

D

500

600

800

1000

External size(mm)

Exclude  protruding parts

W

1000

1000

1100

1500

H

1700

1850

1950

2100

D

1050

1150

1350

1550

Refrigerant

 R404/R23( when temperature at -70℃)

Refrigeration method

Mechanical compression refrigeration (air cooled / water cooled)

Display device

7.0 inch,TFT multi-colour LCD display device

Power

3phase AC220V±10%  60HZ±0.5

Max power(not all the powers(KW)

6

7

10

6

7

10

7.5

8.5

10

13.8

15.4

18

Max current(A)

27.27

32

36.7

27.27

25

36.7

29.0

25

36.7

36.3

43

50

Temperature range -40~150℃
Temperature fluctuation ≤±0.5℃
Temperature deviation ±2.0℃
Humidity range 20%RH~98%RH
Humidity fluctuation ≤±3%RH
Humidity deviation ±3.0%RH(>75%): ±5.0%RH(≤75%)
Inner volume 14.4cf

Write a review

Please login or register to review

0 reviews / Write a review

{{ selectSpec[item.code]?.label }}

408L Lab Temperature Humidity Chamber Dynamic Climate Chamber 14.4 cubic ft. Energy-saving about 40~50%

  • Model #: AEP851XG
  • Availability: In Stock
  • Before: $23,880.00
  • Now: $15,000.00